Abstract
The Schottky barrier line-scanning X- and γ-ray detector is presented, based on bulk undoped semi-insulating (SI) GaAs suitable for radiography imaging with a high position resolution (the lowest pitch 0.125mm). Basic electrical and detection characteristics of the detector (I–V dependences, pulse height spectra of 241Am gamma source) together with evaluated parameters are demonstrated. The detector is irradiated from the edge with an estimated active area of the strip input cross-section (0.125–0.25)×0.25mm2 and absorption length 1.25 or 2.5mm. A photon-counting readout electronics with 480 channels has been developed, giving a total length of the scanner line of about 12cm, using detectors with pitch 0.25mm. The readout system is based on a 16 channel VLSI chip or on 24 channel modules produced with discrete components in SMD technology. Overall detection efficiency of the developed strip SI GaAs radiation detector reaches a value higher than 50% of the theoretical one (∼75%) for detection of 60keV photons. The status of the development of a modular portable X-ray scanner based on the developed detectors is also presented.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.