Abstract

Aim: To further enhance the detection sensitivity and increase resolving power of top-down intact protein bioanalysis, middle-down approach was explored. Materials & methods: An monoclonal antibody (mAb) was used as a model protein to evaluate quantitative bioanalytical assay performance and a disulfide linked dimer protein was investigated for its pharmacokinetics properties and catabolism in vivo by middle-down approach. Results & Conclusion: For quantitation of the mAb, different subunits generated by middle-down approach provided different level of signal improvement in biological samples with Lc and half Fc giving five-times better sensitivity than intact mAb. For the dimer protein, middle-down analysis by reduction enabled effective differentiation of the unchanged protein and its oxidized form, and clearly elucidated their respective proteolytic catabolites.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.