Abstract

It is proposed to use iron whiskers as elements for magnetic thermometers. The sharpness and reproducibility of the large second harmonic response that occurs just as the centre of the whisker saturates depends upon temperature through Ms(T). As a thermometer the whisker has a built-in fixed point which is reproducible to parts in 106. A one point calibration is needed to determined the demagnetizing factor. Suggestions are made for incorporating magnetic thermometers into thermocouples and resistance thermometers.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.