The radioluminescence (RL) emitted by LiMgPO4 detector under proton beam irradiation was investigated in real time at the radiotherapy facility in the Institute of Nuclear Physics Polish Academy of Sciences in Krakow. The facility uses protons accelerated by the AIC-144 isochronous cyclotron up to the energy of 60MeV. The measurements of RL were carried out using a remote optical fiber device with a luminophore detector and photomultiplier located at opposite ends of the optical fiber. A thin slice of LiMgPO4 doped with Tm (1.2mol%) crystal was exposed to the proton beam. The tested detector allowed for the measurement of proton beam current, flux fluence and determination of proton beam time structure parameters. The investigation of LiMgPO4 crystal showed its high sensitivity, fast reaction time to irradiation and possibility of application as the detector for control of proton beam parameters.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.