Abstract

This work describes the construction and evaluation of lactate sol–gel biosensors to accomplish the determination of lactate in pharmaceutical products. Lactate oxidase was incorporated in a porous sol–gel film placed onto a platinum-based electrode. Acid and basic catalysis were assessed. When coupled to a sequential injection system (SIA) the biosensor, based on (3-aminopropyl)trimethoxysilane, 2-(3,4-epoxycyclohexyl)ethyl-trimethoxysilane, deionised water, polyethylene glycol 6000 and acid catalyst, presented a range of linearity of 5 × 10 −5 to 5 × 10 −3 M. The analytical usefulness of the developed biosensor was evaluated through analysis of commercial pharmaceutical products containing lactate with a sampling rate of 40 samples h −1. The enzyme remained active for at least 30 days, enabling about 700 determinations without sensitivity decrease.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.