Abstract

An ionic liquid–TiO2 nanoparticle modified carbon paste electrode (IL–TiO2/CPE) was used as a fast and sensitive tool for the investigation of the electrochemical oxidation of benserazide using voltammetry. This modified electrode has been fabricated using hydrophilic ionic liquid (n-hexyl-3-methylimidazolium hexafluoro phosphate) as a binder. The modified electrode offers a considerable improvement in voltammetric sensitivity toward benserazide, compared to the bare electrode. Using differential pulse voltammetry (DPV), the electrocatalytic oxidation peak current of benserazide shows a linear calibration curve in the range of 1.0–600μmolL−1 benserazide. The limit of detection was equal to 0.4μmolL−1. The relative standard deviation (RSD%) for eight successive assays of 10μmolL−1 benserazide was 1.1%. Finally, the proposed method was successfully applied to the determination of benserazide in real samples such as blood serum and urine.