Two new simple and effective methods based on dispersive liquid-liquid microextraction (DLLME) procedure, termed "in-syringe DLLME (IS-DLLME)" and "narrow-bore tube DLLME (NB-DLLME)", were developed and applied for rapid and simultaneous separation and preconcentration of trace amounts of benzene, toluene, ethylbenzene and xylene isomers in water samples followed by gas chromatographic analysis. Different parameters influencing the extraction efficiency of both methods such as type and volume of the extraction solvent and the disperser solvent; pH, temperature and volume of sample solution and ionic strength of samples were investigated and optimized. Under optimal condition, the limits of detection ranged from 1.7 to 2.4 µg L(-1) for IS-DLLME and 1.5 to 2.2 µg L(-1) for NB-DLLME. Precision (as relative standard deviation) of the two techniques was between 2.1 and 4.6% for IS-DLLME and between 1.5 and 4.5% for NB-DLLME. The enrichment factors found to be between 20-29 and 31-73 for IS- and NB-DLLME, respectively. The applicability of the proposed methods was investigated by analyzing real water samples.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.