Abstract
The acidity of several types of catalyst was measured by means of in situ TG, using the procedure of NH 3 adsorption—desorption, with a modified quartz furnace tube on a Du Pont 9900 Computer/Thermal Analysis System. HY and HZSM-5 zeolites were used as reference samples to test the repeatability and/or precision of this method. The total acid amount and acid strength of samples could be directly calculated from the mass change on TG—DTG curves of NH 3 adsorption—desorption. The relative standard deviation of the data which were collected from five measurements of HY is less than ±0.050. This result was examined using HZSM-5 and the data were comparable with those reported in the literature. The relations between the acidities and the catalytic activities for Pt/MgO-cross linked montmorillonite (Pt/MgO-CLM) and β zeolite catalysts are also discussed.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.