Abstract
Triadimenol is the most important triazole fungicide used for prevention and treatment of plant fungal disease. A simple high performance liquid chromatography (HPLC) method for triadimenol analysis in water sample is reported in this study. The optimized HPLC system using C18 column was achieved using mobile phase composition containing acetonitrile-water (70:30, v/v), a flow rate of 1.0 mL/min, UV detection at 220 nm, and analysis time within 2.5 min. The calibration graph was linear in the range 5 to 50 mg/L with r2 of 0.9997. The limit of detection (LOD) and limit of quantification (LOQ) obtained were 1.23 and 4.12 mg/L, respectively. Combination of solid phase extraction (SPE) and the optimized HPLC method has been successfully applied for the determination of triadimenol in water sample with a recovery of 94.24 % (RSD = 5.38%, n = 4). The proposed HPLC method provides short analysis time, high reproducibility and high sensitivity.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.