Abstract
As an alterative to photodiode array or mass spectral analysis, a heart-cutting two-dimensional liquid chromatography technique (2D-LC) can be utilized to assess peak purity. In this work, the strengths of the heart-cutting 2D-LC technique are presented by highlighting a case study where a co-eluting impurity present at 0.8 % was non-detectable using traditional methods. However, when fractions taken across the main peak were re-assayed using an orthogonal method, this peak was readily observed. With modern switching valves and up-to-date chromatography software, implementing this technique into a validation protocol for assessing peak purity is simple and can be accomplished through only minor modifications to existing HPLC equipment.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.