Abstract

As an alterative to photodiode array or mass spectral analysis, a heart-cutting two-dimensional liquid chromatography technique (2D-LC) can be utilized to assess peak purity. In this work, the strengths of the heart-cutting 2D-LC technique are presented by highlighting a case study where a co-eluting impurity present at 0.8 % was non-detectable using traditional methods. However, when fractions taken across the main peak were re-assayed using an orthogonal method, this peak was readily observed. With modern switching valves and up-to-date chromatography software, implementing this technique into a validation protocol for assessing peak purity is simple and can be accomplished through only minor modifications to existing HPLC equipment.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.