Abstract

In Electrospray Droplet Impact (EDI), the charged liquid droplets formed by ambient electrospray are introduced through an orifice into vacuum, accelerated by 10 kV and impact the samples prepared on the metal substrate. The secondary ions are detected by an orthogonal-type time-of-flight mass spectrometer. EDI can afford soft ionization/desorption for various kinds of real-world samples such as biological tissues, dyes, pigments, synthetic polymers, and inorganic materials with no special sample preparation. EDI is capable of atomic- and molecular-level etching with leaving little damage on the etched surface. Moreover, the useful yield (total ions generated divided by the total atoms or molecules desorbed) was found to be larger than 10-2. Due to these unique natures, EDI may be promising for the next-generation nano-scale 3D imaging.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.