Abstract

Cholesterol oxidase (ChOx) has been immobilized onto electrochemically prepared poly- N-methylpyrrole P(NMPY)- p-toluene sulphonate (PTS)/indium–tin-oxide (ITO) films by physical adsorption and entrapment techniques, respectively. The ChOx(entrapped)-P(NMPY)-PTS/ITO, ChOx(physisorbed)-P(NMPY)-PTS/ITO along with ferrocene carboxylic acid (FeMC) mediated ChOx(physisorbed)-P(NMPY)-PTS/ITO films have been characterized using cyclic voltammetry (CV), Fourier-transform-infrared (FTIR) spectroscopy and scanning electron microscopy (SEM) techniques, respectively. Compared to ChOx(entrapped)-P(NMPY)-PTS/ITO (0.205 mA (mM cm 2) −1) and ChOx(physisorbed)-P(NMPY)-PTS/ITO (0.24 mA (mM cm 2) −1) electrodes, the ChOx(physisorbed)–FeMC(physisorbed)-P(NMPY)-PTS/ITO bio-electrode exhibits increased sensitivity (0.45 mA (mM cm 2) −1) and decreased response to interferent (ascorbic acid, uric acid, urea, glucose). A tentative mechanism for the immobilization of cholesterol oxidase onto P(NMPY)- p-toluene sulphonate (PTS) films has been proposed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.