AbstractAn anodic stripping voltammetric procedure for the determination of bismuth in the presence of excess of Cu2+ ions at two ex situ plated gold film electrodes was described. The procedure is based on utilization of two deposition and two stripping steps system. The presented procedure ensures increasing the sensitivity of Bi3+ determination and minimization of interferences related to peaks’ overlapping. The calibration graph for bismuth determination was linear from 2.5×10−9 to 2×10−8 mol L−1 for deposition time of 300 s at both working electrodes while detection limit was 7.7×10−10 mol L−1.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.