Abstract
Coronary stents are mesh like structures having complex geometrical shapes. Their complex geometry makes it difficult to estimate their surface area exactly which is necessary in different experiments to study the degradation behavior. There are methods available to calculate or estimate the stent outer surface area which is also called Metallic Surface Area (MSA). In this vicinity, no method available in the literature to determine the total surface area of stent with all exposed sides. In this work, a non-destructive method is presented to calculate total surface area of stents by using micro computed tomography (μCT) measurements. Since the corrosion rate calculations need exposed surface area of samples. Therefore, with the help of this method, stents can be used as a test samples instead of regular geometrical shaped samples for corrosion tests.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.