Abstract

A laboratory hard X‐ray photoelectron spectroscopy (HXPS) system at 5.4‐keV excitation energy was used to measure the angle dependence of a silicon oxide overlayer on a Si(0 0 1) substrate with overlayer thickness ranging from 4 to 25 nm. The thickness values of the SiO2 overlayers were determined by utilizing a focused monochromatized Cr Kα source and a high‐energy hemispherical analyzer with an angle‐resolved wide acceptance angle objective lens. The modulation of the photoemission intensity due to photoelectron diffraction, which deteriorates high‐precision thickness determination, was suppressed significantly by continuous sample rotation around the sample's normal during the measurements. The resultant thickness values very well agree with those determined by ellipsometry in the same sample set. To demonstrate merits of the large information depth measurements, profiling of a wedged SiO2 layer buried in a gate stack model structure with Ir (8 nm) and HfO2 (2 nm) overlayers was performed. Copyright © 2011 John Wiley & Sons, Ltd.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.