Abstract

A commercially available fenitrothion-ELISA kit, designed for residue analysis of crop samples, was applied in a method development study using 10 different soils spiked with the target pesticide. Recoveries determined by gas chromatography (GC) were compared to those in ELISA analysis. Recoveries in ELISA were biased high in six soil samples which had low pH and high sand content. The range of recoveries in 10 soils was from 87 to 163% while, in contrast, GC recovery was 72-86%. Soil matrix, such as high-molecular-weight organics and divalent cations, influenced the ELISA reaction to cause an overestimation of recovery.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.