An array of calorimetric low temperature detectors (CLTD’s) for energy sensitive detection of heavy ions was combined with time-of-flight (TOF) detectors to obtain a detector system for high resolution mass identification of low energy heavy ions. In addition the same setup was used to prove the ability of CLTD’s to be used in electronic stopping power measurements for heavy ions in matter. Experiments with 50 MeV 63Cu and 65Cu ions at the tandem accelerator at the MPI at Heidelberg, and with 25 to 250 MeV 238U ions at the UNILAC accelerator at GSI at Darmstadt have been performed. For 63,65Cu at 50 MeV a mass resolution of Δm(FWHM)=0.9 amu, and for 238U in an energy range of 65 to 150 MeV a resolution of Δm(FWHM)=1.28 amu, was obtained. The results for stopping powers of 238U in carbon and gold are presented and compared with theoretical predictions and data from the literature.