Abstract

Objective To apply biological variation and six Sigma models to evaluate analysis performance of 6 HbA1c analyzers and design the new quality control strategy. Method We collected data of imprecision and inaccuracy from routine internal quality control (June 2017–December 2017) and proficiency test of NGSP, respectively. The coefficient of variance (CV)% and bias% were plotted in the biological variation and six sigma models. The new quality control strategy was designed by the sigma value and OPSpecs. The quality improvement was guided by the QGI. Results The analytical performance of 6 HbA1c analyzers in our laboratory were good in the routine model, However, 50% (3/6) and 67% (4/6) of the HbA1c analyzers reached the acceptable level in the biological variation and six Sigma model, respectively. We also design personalized control strategy and promote quality improvement by combining the sigma value, OPSpecs, and QGI. Conclusions Biological variation model and six sigma model could visually display the performance of 6 HbA1c analyzers and personalized control strategy could be designed based on the sigma value, OPSpecs, and QGI.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.