Abstract

When determining element contents in hair samples without sample digestion it is necessary to analyze large sample volumes in order to minimize problems of inhomogeneity of biological sample materials. Therefore an electrothermal vaporization system (ETV) is used for solid sample introduction into an inductively coupled plasma (ICP) for the determination of matrix and trace elements in hair. This paper concentrates on the instrumental aspects without time consuming sample preparation. The results obtained for optimization tests, ETV operating parameters and ICP operating parameters, are shown and discussed. Standard additions are used for calibration for the determination of Zn, Mg, and Mn in human hair. Studies including reproducibility and detection limits for chosen elements have been carried out on certified reference materials (CRMs). The determination of reproducibility (relative standard deviation (RSD) of n = 10) and detection limits (DLs) of Zn (RSD < 8.5%, DL < 0.8 μ g −1), Mn (RSD < 14.1%, DL < 0.3 μ g −1), and Mg (RSD < 7.4%, DL < 6.6 μ g −1) are satisfactory. The concentration values found show good agreement with the corresponding certified values. Further sample preparation steps, including hair sampling, washing procedure and homogenization for hair, relating to measurements of real hair samples are described.