Abstract
In pesticide residue analysis, screening for over 150 compounds has to be performed on a daily basis. As part of the quality control measures it is crucial to verify that the chromatographic system fits the purpose, or if any deterioration occurred during its previous use. The operation conditions of the chromatographic system can be best monitored with properly selected system suitability test (SST) mixtures, which provide information with one injection on the characteristic performance parameters of the whole system from the injector to the detectors. We developed SST mixtures that are also suitable for use with electron-capture, nitrogen–phosphorus and pulse flame photometric detectors. These SST mixtures were applied over 3 years to monitor the system performance parameters, such as the number of effective theoretical plates, resolution, asymmetry, detection limit and selectivity. The applicability and advantages of these tests are illustrated and discussed.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.