Abstract

A portable total reflection X-ray fluorescence (TXRF) spectrometer has been applied to a trace elemental analysis of wines. Sulfur, K, Mn, Fe, and Rb were detected. These five elements were quantified by using 1 ppm of Co as an internal standard. The quantified concentrations ranged from sub-ppm to several hundred ppm. Because of organic substances in wines, the scattering of the incident X-rays from the dry residues of wines becomes strong. Therefore, a high spectral background appears in TXRF spectra of wines. Because of this background, relative standard deviations of the quantified concentrations were from 4 to 28%. Although the high spectral background appeared in the TXRF spectra of the wines, a detection limit down to several tens of ppb was achieved. The present portable spectrometer can be applied to screening for trace elements in wines before an accurate and precise analysis using a large elemental analyzer.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.