Abstract
In this work, a water-soluble polymer, polyethylenimine (PEI) was used for the simultaneous separation and preconcentration of trace Cu and Mn prior to their determination by flame atomic absorption spectrometry. For this purpose, the sample and the PEI solution were mixed and the metal-bound polymer was precipitated by adding acetone. The precipitate was separated and dissolved in a minimum amounts of water and aspirated into a flame AAS. By increasing the ratio of the volumes of sample to water used in dissolving the precipitate, the analyte elements were concentrated as needed. The sorption is quantitative in the pH ≥6. Detection limits were 5.2 μg/L for Cu and 5.4 μg/L for Mn. This method is simple, fast and precise.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.