Abstract

Raman spectroscopy is a highly specific technique for the identification of molecules by way of the associated characteristic spectra. The aim of this feasibility study is to assess the combination of the multivariate calibration technique of Partial Least-Squares with Raman spectroscopy for the estimation of glucose, lactic acid, and urea concentrations in the presence of each other in a water substrate. The instrument is a CCD-based Raman spectrometer utilizing the 514.5 nm argon laser line. The estimates for the analyte concentrations yielded a standard deviation of concentration residuals of 20.71 mg/dL for glucose, 12.92 mg/dL for lactic acid, and 19.07 mg/dL for urea.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.