A fiber bundle has been designed for measurement of gas concentrations in a chemical vapor deposition (CVD) chamber. The bundle is broadband, covering the spectrum range from 500 to 6000 cm(-1). Using this bundle, we have designed Fourier Transform infrared (FTIR) sampling configurations with reduced beam divergence (from 0.4 to 0.1 rad) and focal spot diameters (from 12 to 4 mm). The FTIR spectrum of tetraethoxysilane vapors in a CVD chamber was determined with the bundle and modified SAM connectors. An ellipsoid reflector was designed that yielded improved signal-to-noise ratios (SNR's) in the fiber-based spectra by a factor of 2. The incorporation of this reflector in spectrometers by use of the fiber bundle has the potential to reduce the sampling times for spectra while retaining acceptable SNR's.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.