Abstract

A series of halogenated and oxygenated heterocyclic, aromatic and aliphatic compounds were evaluated for elemental calibration with the atomic emission detector for gas chromatography (GC–AED). The GC–AED monitored C, Cl, F, N and O signals for these compounds and elemental response factors (ERF) were evaluated for compound independence. The variation in ERF for C, Cl, F, N and O were evaluated and found to be less than 3, 6, 5, 6 and 5% relative standard deviation (RSD), respectively, for all the compounds tested. Significant ERF compound dependence was identified for Cl in one of the compound groups and F in all compound groups. However, the ERF variation only increased from 5 to 6% RSD for Cl and from 2 to 5% RSD for F after including the compound dependent bias. Compound independent calibration can be utilized if these errors are acceptable for a particular application. The linearity of ERF with concentration and GC-AED reagent gas composition were also evaluated.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.