Abstract

A commercially available sample-injection valve (manual version) has been coupled to the nebulizer of a premixed laminar-flow burner on an atomic absorption spectrometer for metal analysis in volumes of 50–500μl. For 500-μl samples statistical analysis showed an overall coefficient of variation of 1.7% for readings from the peak height facility of a digital read-out unit. In order to obtain reproducible results with 50-μl samples (overall coefficient of variation of 4.3%) it was necessary to use the peak height data obtained from a pen-recorder connected to the digital read-out unit.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.