Sulphur-containing gases are converted by non-catalytic reductive pyrolysis into an equivalent amount of H 2S, which is determined by electrochemical oxidation at a gold-plated porous membrane electrode. The calibration for H 2S is linear up to 18 ml/m 3 with a slope of 1.12 μA.m 3.ml −1. The relative standard deviation for the determination of 3.7 ml/m 3 H 2S was 1.0% and the detection limit 4 μl/m. About 60 samples can be determined per hour. The method has been employed for the determination of total sulphur and individual gases after separation on a chromatographic column.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.