Abstract

Recent developments in amplifier hardware enable low‐noise measurements of exceedingly small ion beams in isotope ratio analysis, yielding higher precision from smaller samples than ever before. To date, these amplifiers have largely been employed in thermal ionisation instruments, with few applications using plasma‐source (i.e., MC‐ICP‐MS) instruments. Here, we demonstrate the utility of these new generation 1013 Ω amplifiers in MC‐ICP‐MS, employing Pt isotopes as a case study, a system that could greatly benefit from the promised advances. The data demonstrate that for samples with low Pt abundance, for a modest increase in uncertainties, the amount of sample required can be reduced by a factor of 50–100. This technique thereby opens up new possibilities for analysis of samples that have low Pt mass fractions or were otherwise impossible to obtain in sufficient quantities.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.