Abstract

Objective To observe the clinical application and nursing of Icare rebound tonometer.Methods Intraocular pressure (IOP) measurements were performed on 112 cases (112 eyes) with ICARE rebound tonometer, measure on central cornea 2mm distance to temporal and nasal limbus corneae respectively.Results The results of intraocular pressure measurement on central cornea, 2mm distance to temporal and nasal limbus corneae are (15.4±3.4),(15.8±3.6),(15.6±3.4)mm Hg. There were no significant difference among them(F=0.325,P>0.05), but they have a good consistency.Conclusions There was good consistency on three different position on IOP measurements. The nursing is important for correct measurement. Key words: Tonometry,ocular; Nursing; Icare rebound tonometer

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.