Abstract

Palladium (Pd) is extensively used in pharmaceutical small molecule drug substance processes, however it must be removed prior to release of the active pharmaceutical ingredient (API). Evaluation of four TXRF instruments and configurations were compared to ICP-MS instrumentation for trace metal analysis, most importantly for Pd. Standards and six pharmaceutical drug substances, triprolidine HCl, diphenhydramine HCl, chlorpheniramine maleate, pseudoephedrine HCl, ephedrine sulfate, and scopolamine HBr, were analyzed to determine linearity, sensitivity, accuracy, and precision for Pd plus Cr, Fe, Cu, Rh, and Pt versus interferences, particularly from Cl, S, and Ar, on the various X-ray fluorescence lines. Irrespective of instrument platform, in general X-ray sources capable of accessing Pd-K lines were found to be most effective in determination of Pd in APIs.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.