Abstract

AbstractSeveral methods for the routine determination of crystallite size by means of X-ray diffraction line-broadening have previously been reported. Although these techniques have proven useful and reliable when utilized with the single X-ray diffractometer and instrumental geometry used to originally develop the methods, it was not known whether other instruments would provide similar reliability. Therefore a study was performed to evaluate the applicability of routine methods of crystallite size analysis to other X-ray diffraction units. A series of six beryllium oxide powder specimens, whose average crystallite sizes ranged stepwise from about 35 to nearly 3000 Å, were used to test a number of X-ray diffractometers. By using a predetermined diffraction geometry for each instrument tested, measured crystallite sizes were found to be quite reproducible and well within the limits of experimental error. The testing procedure, instrumental conditions, and individual performance results are presented in this paper.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.