Abstract
To investigate the effects of intense radiation on the operation of the liquid xenon spectrometer we have created apparatus complex on the basis of the liquid xenon detector. The experimental setup consists of a multifunctional chamber, gas system, cooling system, temperature control system, X-ray generator, a special preamp, passive protection, scintillation monitor of the accelerator beam, thermoluminescent dosimeters, copper monitor bremsstrahlung, Ge(Li) detector. Multifunctional chamber includes a detecting unit (flat or cylindrical ionization chamber), the cleaning unit of the xenon, control unit of the purity of liquid xenon. The liquid xenon detector was irradiated by bremsstrahlung pulses of the microtron. The frequency of irradiation pulses was 400 Hz. The absorbed dose was varied from 10-7 to 0.1 Gy per pulse. The electronic and ionic processes in liquid xenon at different radiation doses were investigated. The recovery time of the spectrometric mode of operation of the liquid xenon detector after intense pulse irradiation has been studied. Stable operation of the liquid xenon spectrometer in the intervals between the pulses of the accelerator shown for a long time.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.