Abstract

Drying instrumented canals with pressurized air may result in patient morbidity or even fatality. Low pressure and side vent needles have been suggested to lessen the danger. This study observed apical pressures from different needles inserted deeply into small round and ovoid canals as instrumentation progressed. Low-pressure (5 psi) air was injected through the needles, and apical pressures were recorded after each instrument. Pressures varied greatly within each test group. Generalities that can be drawn are that binding the needle within the canal gives higher pressures than with the needle slightly short of binding and that pressures were higher with apexes instrumented to size 30 and higher. With the needle tightly bound, neither needle size, needle design, nor canal shape resulted in statistically significant mean pressure differences. With the needle slightly withdrawn, larger bore needles gave higher pressures than small diameter needles. Caution is advised with the clinical use of pressurized air in the drying of root canals.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.