Abstract
For linear light scattering in apertureless scanning near-field optical microscopy, we have studied the correlations between the tip radius of the probe, signal strength, spatial resolution, and sample material. Pronounced variations of the near-field distance dependence on tip shape and dielectric function of the sample are observed. For very sharp metal tips, the scattered near-field signal decays on a 5 nm length scale. Despite this highly localized tip–sample coupling, the contrast is found to depend sensitively on the vertical composition of the sample on a length scale given by the penetration depth of the incident light. The resulting implications on the use of the technique as an analytic probe method are discussed.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.