Abstract

We analyzed 50 aortic valve replacements (AVR) with Kemerovo-AB-Neo stentless bioprostheses. Mean age was 66.75 (54 to78) years, hospital mortality rate accounted for 2%. The peak transprosthetic pressure gradient (PTPG) in patients operated for aortic stenosis came to 19.57 mm Hg. PTPG was shown to be dependent on the implantation technique; it reached 24.57 mm Hg when in order to fix the proximal line, interrupted sutures were used and ran to 175 mm Hg in the case of running sutures (р = 0.03). Helical CT confirmed fine mobility of the cardiac structures after Kemerovo-AB-Neo implantation: the aortic cross-section area varied up to 84% during the cardiac cycle. It should be noted that AVR with Kemerovo-AB-Neo stentless valves gives good clinical and hemodynamic results early after surgery. A free-hand technique of implantation should be preferred. The implantation of stentless bioprostheses retains cardiac structures mobility and natural aortic root dimensions after performing an AVR.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.