Abstract

Stent migration is a significant obstacle to successful stent placement. There has been no investigation of the effect and quantitative interpretation of flaps attached to a plastic stent (PS) on antimigration. The antimigration effects of the number of flaps on a PS in a 3D-printed pancreatic phantom (3DP) and extracted porcine pancreas (EPP) were investigated. Four PS types were used in this study: stent without flaps (type 1), stent with two flaps (type 2), stent with four horizontally made flaps (type 3), and stent with four vertically made flaps (type 4). The stents were measured and compared for antimigration force (AMF) in the 3DP and EPP using a customized measuring method and an integrated measuring device. The mean maximum AMFs (MAMFs) in types 2, 3, and 4 were significantly higher than that in type 1 (all p < 0.001). Moreover, the mean MAMFs in types 3 and 4 were significantly higher than that in type 2 (all p < 0.001). When the flaps were removed from the pancreatic duct, the AMF decreased rapidly. As the number of flaps increased, the antimigration effects significantly increased in the 3DP and EPP. However, the direction of the flaps did not affect the MAMF. The position of the flaps attached to the surface of the stent affected the AMF.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.