Abstract

The incidence of catheter associated urinary tract infections (CAUTIs) is increasing worldwide. This study was designed to modify a biomaterial by impregnating a silicone urinary catheter with combination of a macrolide, azithromycin (AZM) and a fluoroquinolone, ciprofloxacin (CIP). Drug release profiles showed slow yet continuous release of antibiotics from catheters for one month. In vitro efficacy testing showed that group B catheters [3% (w v−1) CIP + 6% (w v−1) AZM] outperformed group A catheters [2% (w v−1) CIP + 5% (w v−1) AZM] by (1) showing larger zones of inhibition (>31 mm) compared to group A (<28 mm) for up to 30 days against Pseudomonas aeruginosa PAO1; (2) killing adhered bacteria in 24 h compared to 24–48 h in group A; (3) showing longer antimicrobial durability for four weeks; and (4) exhibiting a stable real-time shelf life of one year, suggesting that these catheters can be explored in clinical settings, especially in long-term CAUTI.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.