Abstract

BackgroundCovering insertion sites with chlorhexidine impregnated dressings has been proven to be clinically effective in reducing catheter related blood stream infections (CR-BSI). Two chlorhexidine gluconate (CHG)-impregnated dressings are commercially available, a polyurethane foam disk and a film dressing containing a chlorhexidine gluconate-impregnated gel pad. While both have demonstrated efficacy in clinical settings, the major drawback of high cost and impaired IV insertion site visibility limits their usage. A new, simple film dressing containing CHG within its adhesive layer is now available. The objective of this study was to test the in vitro antimicrobial efficacy of the new dressing in comparison to the CHG-impregnated gel dressing.MethodsQuantitative aliquots of suspensions (concentration of 1.0x106 to 5.0x106 cfu/sample) of clinically relevant challenge organisms (Staphylococcus species, gram-negative bacilli, Candida albicans) were incubated in contact with the new CHG-containing film dressing, a placebo version of the same (negative control) and the commercially available CHG-impregnated gel dressing (positive control). Serial dilutions of the surviving organisms were quantified using the pour plate after 1, 3, 5, and 7 days of incubation in order to calculate an antimicrobial log10 reduction for each organism/dressing combination at each point in time.ResultsThe new CHG-containing film dressing delivered greater than 5.0 log10 reduction throughout the 7 days on all aerobic gram-negative bacilli and Staphylococcus species tested. As of day 1 the CHG-containing film dressing provided greater than 5.0 log10 reduction on Candida albicans. There were no statistically significant differences in the log10 reduction between the two dressings tested.ConclusionThe new CHG-containing film dressing was found to be as effective as the chlorhexidine gluconate-impregnated gel dressing on clinically relevant microbes.

Highlights

  • Covering insertion sites with chlorhexidine impregnated dressings has been proven to be clinically effective in reducing catheter related blood stream infections (CR-BSI)

  • The new chlorhexidine gluconate (CHG)-containing film dressing delivered greater than 5.0 log10 reduction throughout the 7 days on all aerobic gram-negative bacilli and Staphylococcus species tested

  • As of day 1 the CHG-containing film dressing provided greater than 5.0 log10 reduction on Candida albicans

Read more

Summary

Introduction

Covering insertion sites with chlorhexidine impregnated dressings has been proven to be clinically effective in reducing catheter related blood stream infections (CR-BSI). Two chlorhexidine gluconate (CHG)-impregnated dressings are commercially available, a polyurethane foam disk and a film dressing containing a chlorhexidine gluconate-impregnated gel pad. While both have demonstrated efficacy in clinical settings, the major drawback of high cost and impaired IV insertion site visibility limits their usage. A new, simple film dressing containing CHG within its adhesive layer is available. The objective of this study was to test the in vitro antimicrobial efficacy of the new dressing in comparison to the CHG-impregnated gel dressing

Objectives
Methods
Results
Conclusion

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.