Abstract

We measured the stability of vancomycin, amikacin, netilmicin, tobramycin, gentamicin, clindamycin, and aztreonam in an implantable drug pump. Two pumps were filled with the antibiotic being tested and were incubated at 37 degrees for 3 weeks. Samples were taken daily. At 3 weeks the pumps were disassembled and examined microscopically for signs of deterioration. If the pumps were intact, bioassays were used to quantitate the biologic activity of each daily antibiotic sample. Vancomycin degraded the pump. Amikacin, netilmicin, tobramycin, gentamicin, clindamycin, and aztreonam had no effect on the pump and maintained their biologic activity over the 3-week period.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.