Abstract

Anti-Stokes/Stokes Raman peak intensity ratio was used to infer sample temperatures, but the influence factors of system correction factors were not clear. Non-contact in-situ anti-Stokes/Stokes temperature calibration was carried out for up to 1500 K based on six different samples under two excitation light sources (±50 K within 1000 K, ±100 K above 1000 K), and the system correction factor γ was systematically investigated. The results show that the correction factor γ of anti-Stokes/Stokes thermometry is affected by the wavelength of the excitation light source, Raman mode peak position, temperature measurement region and other factors. The anti-Stokes/Stokes thermometry was applied to the laser-heating diamond anvil cell (LHDAC) experiment to investigate the anharmonic effect of hBN under high temperature and high pressure. It is concluded that the strong anharmonic effect caused by phonon scattering at low pressure gradually changes into the predominance of localized molecular lattice thermal expansion at high pressure.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.