Abstract
The purpose of the study was to determine the accuracy of half-Fourier single-shot rapid acquisition with relaxation enhancement high-spatial-resolution magnetic resonance (MR) imaging performed with a microscopy coil in the diagnosis of narrow anterior chamber angle in patients with glaucoma. Slit-lamp biomicroscopy served as the reference standard. The institutional review board approved this study, and written informed consent was obtained from the 20 recruited patients. There was excellent agreement between MR gonioscopy and slit-lamp biomicroscopy in the classification of anterior chamber angles as narrow or open (kappa = 0.89 [95% confidence interval: 0.69, 1.10]). MR gonioscopy has substantial potential as a technique used to evaluate glaucoma.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.