Abstract

Objective: To assess the correlation between individual computerized cardiotocography (cCTG) variables and acid-base status at birth before the first breathing in uncomplicated term pregnancy delivered by cesarean section. Study design: A retrospective cohort study. Subjects and methods: Seventy singleton normal pregnancies delivered by elective cesarean section. The last antepartum cCTG performed within 4 h from delivery by the System 8002 (Oxford Sonicaid, UK) was correlated to umbilical blood gas analysis (UBGA) values (AVL compact 2 analyser). Considering blood gas analysis values as dependent and individual cCTG parameters as independent variables the best regression subsets followed by a backward stepwise linear regression was used. Results: There was no significant correlation of cCTG parameters with any of the values of blood gas analysis. However, when neonatal conditions, as expressed by arterial pH>7.20, Becf>−4.0 mmol/l and Apgar score at 5 min >7, were taken as an end-point, they could be predicted by Acc15 and FM/hour (sensitivity: 35%, positive predictive value: 92%, cut-off>7 and 67%, 92%, >21, respectively). Conclusion: In uncomplicated pregnancies, the prospect of a vigorous newborn seems particularly associated with the presence of Acc>15 and FM/hour>21.