A graphite (carbon) composite solid electrode, prepared from graphite powder and epoxy resin, was used as a working electrode for anodic stripping voltammetry. The underpotential deposition effect, which appears at metallic electrodes, was clearly observed on this type of electrode as well. In the case of a simultaneous deposition of two metals on the surface of the composite solid electrode, the anodic dissolution of the metal, which is anodically dissolved at more negative potentials, is substantially influenced by the presence of the other deposited metal. This effect was exploited for the determination of lead in the presence of other metals by differential pulse anodic stripping voltammetry. The article presents possible applicability of such a type of very simple composite electrode to studies of the underpotential deposition effect as well as for analytical purposes.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.