Abstract

AbstractGold macrodisk, microdisk, and microband electrodes are compared for the anodic stripping voltammetry of Se(IV). All of these electrodes show a stripping peak at 0.79 V with a width at half height of 80–90 mV. The linear dynamic range, limit of detection, sensitivity, and precision are compared when using the three electrodes. Although flux at the microelectrodes is greater than at the macroelectrode, the limit of detection is similar for all three geometries. The linear dynamic range is smallest when using the microdisk electrode. The precision of the peak current is best with the macrodisk electrode. Although microelectrodes enjoy many advantages in analyses based on reversible systems, these advantages are not always realized with irreversible systems.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.