HPGe detector is used widely, its property is affected by neutron. Annealing to the germanium crystal is a good approach to recover the detector. To raise the temperature of germanium crystal to recover the resolution for the HPGe detector. A special copper ingot is designed and manufactured to increase the heating efficiency. Two temperature controlled power supplies are used to heat the germanium crystal and keep the temperature for cooling rod. The experiment results indicate that raising temperature to 70 $${}^\circ \hbox {C}$$ for five days, the restoration efficiency is better than 90%.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.