The article presents results of biomechanical bench tests of bone fragment fixation using the model of a transverse diaphyseal fracture in a long cortical bone in two mutually perpendicular planes for different modes of installation of flexible titanium rods by the flexible intramedullary nailing (FIN) method. The obtained results revealed a significant impact of the bending curvature of the rods, the plane of their placement, and the diameter and inclination of the perforation hole for rod introduction on the rigidity of fixation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.