Abstract
The angular response of luminescent dosimeters (LD), in particular TLD and OSLD, has been compared by applying 6 MV X-ray photons from Versa HD clinical linear accelerator. The study admitted for the irradiation of TLD (n = 475) and OSLD (n = 475) under phantom set up in various gantry angles from 00 to ±900 and various field sizes from 10 x 10 cm2 to 30 x 30 cm2. The variance in the output was observed between 4.4% for TLD and 3.9% for OSLD. A significant deviation from the desired output was detected, towards the angle of incidents, at ±800 to ±900. Additionally, there is no evidence of variation in the dose measurement due to the difference in field size. These results demonstrate a good approximation to the vendor-specified tolerance limits, justifying the use of these LDs within angular incidents of radiation up to ±700. The TLD and OSLD better dose-response is achieved to a gantry angle up to ±700 from the perpendicular incidents. The result shows that both TLD and OSLD could be used as dosimeters for a treatment field that does not extend beyond ±700 beam angle.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.