Abstract
Tibial component malposition in total knee arthroplasty can cause early failure and result in the need for revision surgery. Excessive posterior slope of the tibial component is an uncommon, but difficult problem to correct. Component revision risks additional loss of deficient posterior tibial bone stock. As an alternative, a custom-made angled insert can correct for component malalignment and possibly avoid prosthesis revision. In this report, we present the use of a custom-angled bearing for sagittal alignment correction with 6-year follow-up. The procedure is simple and has little risk, but the cost, potential limitations for correcting excessive deformity, and uncertainty of long-term results make the use of angled inserts a rare and selective procedure. However, when possible to use, an angled insert may be a useful alternative to complete prosthetic revision.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.