Abstract

PurposeTo compare the consistency of Pentacam Scheimpflug system (Pentacam® HR), ray-tracing aberrometry (iTrace), and swept source optical coherence tomography (IOLMaster 700) measurements for Angle Kappa and angle Alpha. MethodsA prospective randomized cohort study. 86 individuals (86 eyes) aged 19–45 years with best corrected vision of 4.9 or better were randomly selected from January 2022 to December 2022 in a tertiary-level hospital. Angle Kappa and angle Alpha were measured or calculated using Pentacam, iTrace, and IOLMaster 700, respectively. ResultsThe mean difference of angle Kappa between any two instruments was not statistically significant, but the mean difference of angle Alpha between both Pentacam and iTrace, Pentacam and IOLMaster 700, and iTrace and IOLMaster 700 was statistically significant (p value <0.001, 0.003, <0.001). The highest consistency of angle Kappa and the narrowest 95% LoA (-0.20 to 0.21) were found between Pentacam and iTrace according to Bland Altman plots, but between Pentacam and IOLMaster 700 according to Mountain plots. Both Bland Altman plots and Mountain plots showed the highest consistency of angle Alpha and the narrowest 95% LoA (-0.14 to 0.24) between Pentacam and iTrace. ConclusionThe mean angle Kappa among Pentacam® HR, iTrace, and IOLMaster 700 had good agreement, and the value of angle Kappa could be output directly, making it more convenient for clinical application. The measured or calculated angle Alpha had poor agreement, and ophthalmologists could refer to measurements from multiple instruments.