Abstract

The detailed mechanism of early-phase arterial healing after novel fluoropolymer-based paclitaxel-eluting stent (PES) implantation in the femoropopliteal (FP) lesions has not been elucidated.Methods and Results:We evaluated the intravascular status of 20 PES implanted in 11 FP lesions of 9 patients using angioscopy at approximately 3 months after implantation. Angioscopic images were analyzed to determine (1) the dominant degree of neointimal coverage (NIC) over the stent; (2) the extent of uncovered struts; and (3) the presence of intrastent thrombus. NIC was classified into 4 grades: grade 0, stent struts fully visible; grade 1, stent struts bulging into the lumen although covered; grade 2, stent struts embedded in neointima, but translucently visible; grade 3, stent struts fully embedded and invisible. The extent of uncovered struts was scored as follows: score 0, no uncovered struts of the entire stent; score 1, uncovered struts area approximately <30% of the stent; and score 2, uncovered struts area approximately ≥30% of the stent. In total, 90% of stents demonstrated grade 1 dominant NIC and 10% showed grade 2 dominant NIC; 85% of stents showed an uncovered stent score of 0, and the remainder had a score of 1. Thrombus was observed in all stents. Widely uncovered stent struts were not observed by angioscopy at 3 months after PES implantation in these FP lesions, even with the detection of thrombus adhesion.